  • Multi parameter atmospheric environment online detection instrument

  • Detection of PM2.5 particle concentration in indoor environment and public open areas; At the same time, it can be used to detect multiple parameters such as temperature, humidity, air pressure, VOC, CO2, O2, formaldehyde, or carbon monoxide in the environment, and its function is very powerful.


Product manual

SPECIFICATION

product name:

Multi parameter atmospheric environment online detection instrument

Model:

INSXX-S-P

version:

V1.0

Creator:

He Genwen

reviewed by:

Yuan Chao

 

version

change content

Changed by

change date

V1.0

new-built

He Genwen

2020-4-24

 

Instrument Usage

Detection of PM2.5 particle concentration in indoor environment and public open areas; At the same time, it can be used to detect various other parameters of the environment, such as temperature, humidity, air pressure, VOC, CO2, O2, formaldehyde, or carbon monoxide, with very powerful functions。

Main working parameters of the comprehensive air quality detector:

Working temperature range

-10 ~ 50℃

Operating humidity range

0 ~ 95%RH

Storage temperature range

-20 ~ 60℃

power input

DC 12~24V

current consumption

The average value is 350mA, and the peak value is 600mA

OUTPUT

Modbus RS4859600bps, 1 start bit, 8 data bits, 1 stop bit, no checksum (optional wifi, ZIGBEE, TCP/IP, lorawan)

size

450g/133mm×133mm×39mm

 

The testing capability and scope of the comprehensive air quality detector

(Customers can choose different parameter combinations according to their own usage requirements):

Main detection parameters

detection range

check the accuracy

Testing methods, etc

PM2.5 Particle concentration measurement

0~500µg/m3

﹤ ±15µg/m3+10%reading

laser scattering

PM10 Particle concentration measurement

0~500µg/m3

﹤ ±15µg/m3+10%reading

laser scattering

Formaldehyde gas

0~2.00ppm;

﹤ ±5% FS

electrochemistry

CO gas

0~500ppm;

﹤ ±5% FS

electrochemistry

Volatile odor detection

(VOC gas)

0~100ppm

﹤ ±20%

semiconductor

combustible

0~100%lel

﹤ ±10% FS

semiconductor

Oxygen concentration monitoring

0~25.0%VOL

﹤ 2%FS

electrochemistry

CO2 detection

400~5000ppm

±75ppm±5%reading

Infrared non dispersive absorption (NDIR)

temperature measurement

-20 ~ 85℃

±0.5℃

Capacitive type

Humidity measurement

0 ~ 100%RH

±3%RH

Capacitive type

atmospheric pressure

200~1200hPa

±0.1%

MEMS

Scalable electrochemical principle sensors (NO/O3/NO2/NO/SO2/H2S/NH3/VOC/PH3/CL2/HCL/HCN/HF), up to 4 sensors can be selected, and the parameters depend on the selected parameters

 

Installation and wiring

  Do not connect the transmitter to a voltage higher than 24VDC.

  Refer to the product appearance and dimension diagram in Figure 1 to install the equipment。

  Install the transmitter in the location where it needs to be tested, away from the heat source or steam source, and prevent direct sunlight. When wiring a circuit, connect the wires to the wiring terminals to ensure that the wiring is correct and firm.

 

Product dimension diagram (unit: mm)

 

Wiring Instructions:

  • Interface Description:

  • From right to left, numbered 1-7, defined as follows:

  • 1. +12V; 2. GND; 3. RS485A; 4. RS485B; 5. I1/V1; 6. I2/V2; 7. TCP/IP

  • Communication protocol section:

  • In the RTU mode using MODBUS protocol, message sending should start with a pause interval of at least 3.5 characters when using RTU mode. When the first domain (address domain) receives it, each device decodes it to determine whether it is being sent to itself. After the last transmitted character, a pause of at least 3.5 characters marks the end of the message. A new message can start after this pause. The entire message frame must be transmitted as a continuous stream. If there is a pause time of more than 3.5 characters before the completion of the message frame, the receiving device will refresh the incomplete message and assume that the next byte is the address field of a new message.

  • 1. The hardware adopts RS485, master-slave half duplex communication, where the host calls the slave address and the slave responds for communication.

  • 2. Data frame format: 1 start bit, 8 data bits, 1 stop bit, no checksum bit.

  • 3. Baud rate: 9600bps.

  • 4. The message frame includes: address domain, function code domain, data domain, and CRC16 detection domain.

  • 5. This device supports hexadecimal 03H and 06H function codes (i.e. decimal 03 and 06 function codes). 03H is used to read the device stack, and 06H is used to modify the device stack.

The corresponding stack plan for the equipment is as follows:

Register address

Corresponding parameters

decimal point

unit

Data Description

40001

CO2

0

ppm


40002

PM2.5

0

ug/m3


40003

O2

1

VOL%


40004

VOC

2

ppm


40005

SMOKE

0

ppm


40006

temperature

1



40007

humidity

1



40008

atmospheric pressure

1

hPa  (hPa)


40009

formaldehyde

2

ppm


40010

carbon monoxide

0

ppm


40011

illumination

0

lux


40012

CH4

0

ppm


40013

noise

1

db


40014

PM10

0

ug/m3


40015





40016

Device Address

0


Default to 1

40017

BAUD

0


default 9600

 

 

Precautions for use:

(1) The RS485 communication line requires shielded twisted pair cables, and if necessary, two 250 Ω terminals with matching resistors are required.

(2) The usage environment should be free of conductive dust, corrosive metals, and gases that damage insulation, avoiding direct sunlight and rain.

(3) Working for a long time in high humidity environments above 90% RH will cause drift.

(4) Avoid placing components in environments with salt spray, acidic or oxidizing gases (such as sulfur dioxide, hydrochloric acid, etc.).

After sales service and warranty terms:

1. Customers can enjoy a one-year warranty service from the date of purchase.

2. The warranty period is calculated from the date of invoice issuance or the date of registration and purchase of the warranty card, and for online sales, it is calculated from the date of receipt.

3. When requesting after-sales service, please contact the relevant person in charge to send it back to our company's after-sales service department.

 

Note: After sales service does not include the following situations:

1. Malfunctions caused by unauthorized disassembly, modification or modification of products;

2. Faults caused by falling during use and handling;

3. The labels on the body must be kept intact, and unauthorized tearing is not guaranteed;

4. Failure to comply with the usage environment of this product resulting in product malfunction;

5. This product is burnt or damaged due to wiring or power supply errors.


  • Website homepage
  • Product Display
  • Qualification certificate
  • Application Information
  • Consulting
  • phone:0755-26588758 15013668479(倪先生) 13922884770(王先生)


    fax:0755-26747274


    mailbox:info@source-sensor.com


    address:East 3rd Floor, Building 15, Majialong Industrial Zone, Nantou Street, Nanshan District, Shenzhen


    Copyright©2002-2023 Yuanjian Sensing Technology . All Rights Reserved Copyright 粤ICP备17008245号-5 technical support:图页网 Website Map xml Website Map html